Initial Performance of a Long Axial FOV PET with TOF and DOI capabilities: IMAS system
Authors
Antonio J. Gonzalez, Alvaro Anreus-Valero, David Sanchez, Santiago Jiménez-Serrano, Marta Freire, Andrea Gonzalez-Montoro, Edwing Y. Ulin-Briseno, Neus Cucarella, John Barrio, Andrew Laing, Jorge Álamo, Julio Barbera, Luis F. Vidal, Marc Gil, Jose M. Benlloch, Alfonso Rios, Luis Marti Bonmati, Irene Torres-Espallardo
Abstract
This work summarizes the design, construction, initial performance evaluation and pilot clinical results of the IMAS system, a long axial field of view (FOV), also known as total-body (TB-), positron emission tomography (PET) prototype scanner. This PET enables for the first time in TB-PET imaging, simultaneously time-of-flight (TOF) and depth-of-interaction (DOI) capabilities. The IMAS detector block is based on LYSO semi monolithic scintillators, with individual slab sizes of 3 mm x 25 mm x 20 mm each. Arrays of 1x8 slabs are coupled to 8x8 Silicon Photomultiplier arrays. A proprietary readout reduces the 64 signals to only 16 outputs, preserving both 3D photon impact positioning and timing accuracy. IMAS has a total of 30,720 channels. PETsys electronics is used for data acquisition. The IMAS geometry is based on 5 rings of 10 cm each, with a 5 cm gap between them. It defines an axial FOV of 71 cm with a bore aperture of 82 cm. We report in this work the pilot tests of the system performance and the first clinical results. We found that the system spatial resolution remained below 4 mm across the entire FOV, even at the off-radial position of 30 cm. A coincidence time resolution with a small size 22Na source of 560 ps FWHM was measured. A sensitivity of 56.54 cps/kBq is in good agreement with previous simulation studies; however, the noise equivalent count rates performance (79 kcps at 3.26 kBq/mL) was significantly lower than expected, likely due to a data transfer bottleneck between the system and the acquisition workstation. Finally, a comparison of one of the imaged patients with a commercial TOF PET/CT scanner is also provided, pinpointing an improved tumor identification for IMAS, and the advantages of TOF and especially DOI capabilities.
Metadata
Related papers
Cosmic Shear in Effective Field Theory at Two-Loop Order: Revisiting $S_8$ in Dark Energy Survey Data
Shi-Fan Chen, Joseph DeRose, Mikhail M. Ivanov, Oliver H. E. Philcox • 2026-03-30
Stop Probing, Start Coding: Why Linear Probes and Sparse Autoencoders Fail at Compositional Generalisation
Vitória Barin Pacela, Shruti Joshi, Isabela Camacho, Simon Lacoste-Julien, Da... • 2026-03-30
SNID-SAGE: A Modern Framework for Interactive Supernova Classification and Spectral Analysis
Fiorenzo Stoppa, Stephen J. Smartt • 2026-03-30
Acoustic-to-articulatory Inversion of the Complete Vocal Tract from RT-MRI with Various Audio Embeddings and Dataset Sizes
Sofiane Azzouz, Pierre-André Vuissoz, Yves Laprie • 2026-03-30
Rotating black hole shadows in metric-affine bumblebee gravity
Jose R. Nascimento, Ana R. M. Oliveira, Albert Yu. Petrov, Paulo J. Porfírio,... • 2026-03-30
Raw Data (Debug)
{
"raw_xml": "<entry>\n <id>http://arxiv.org/abs/2603.09898v1</id>\n <title>Initial Performance of a Long Axial FOV PET with TOF and DOI capabilities: IMAS system</title>\n <updated>2026-03-10T16:52:15Z</updated>\n <link href='https://arxiv.org/abs/2603.09898v1' rel='alternate' type='text/html'/>\n <link href='https://arxiv.org/pdf/2603.09898v1' rel='related' title='pdf' type='application/pdf'/>\n <summary>This work summarizes the design, construction, initial performance evaluation and pilot clinical results of the IMAS system, a long axial field of view (FOV), also known as total-body (TB-), positron emission tomography (PET) prototype scanner. This PET enables for the first time in TB-PET imaging, simultaneously time-of-flight (TOF) and depth-of-interaction (DOI) capabilities. The IMAS detector block is based on LYSO semi monolithic scintillators, with individual slab sizes of 3 mm x 25 mm x 20 mm each. Arrays of 1x8 slabs are coupled to 8x8 Silicon Photomultiplier arrays. A proprietary readout reduces the 64 signals to only 16 outputs, preserving both 3D photon impact positioning and timing accuracy. IMAS has a total of 30,720 channels. PETsys electronics is used for data acquisition. The IMAS geometry is based on 5 rings of 10 cm each, with a 5 cm gap between them. It defines an axial FOV of 71 cm with a bore aperture of 82 cm. We report in this work the pilot tests of the system performance and the first clinical results. We found that the system spatial resolution remained below 4 mm across the entire FOV, even at the off-radial position of 30 cm. A coincidence time resolution with a small size 22Na source of 560 ps FWHM was measured. A sensitivity of 56.54 cps/kBq is in good agreement with previous simulation studies; however, the noise equivalent count rates performance (79 kcps at 3.26 kBq/mL) was significantly lower than expected, likely due to a data transfer bottleneck between the system and the acquisition workstation. Finally, a comparison of one of the imaged patients with a commercial TOF PET/CT scanner is also provided, pinpointing an improved tumor identification for IMAS, and the advantages of TOF and especially DOI capabilities.</summary>\n <category scheme='http://arxiv.org/schemas/atom' term='physics.med-ph'/>\n <published>2026-03-10T16:52:15Z</published>\n <arxiv:primary_category term='physics.med-ph'/>\n <author>\n <name>Antonio J. Gonzalez</name>\n </author>\n <author>\n <name>Alvaro Anreus-Valero</name>\n </author>\n <author>\n <name>David Sanchez</name>\n </author>\n <author>\n <name>Santiago Jiménez-Serrano</name>\n </author>\n <author>\n <name>Marta Freire</name>\n </author>\n <author>\n <name>Andrea Gonzalez-Montoro</name>\n </author>\n <author>\n <name>Edwing Y. Ulin-Briseno</name>\n </author>\n <author>\n <name>Neus Cucarella</name>\n </author>\n <author>\n <name>John Barrio</name>\n </author>\n <author>\n <name>Andrew Laing</name>\n </author>\n <author>\n <name>Jorge Álamo</name>\n </author>\n <author>\n <name>Julio Barbera</name>\n </author>\n <author>\n <name>Luis F. Vidal</name>\n </author>\n <author>\n <name>Marc Gil</name>\n </author>\n <author>\n <name>Jose M. Benlloch</name>\n </author>\n <author>\n <name>Alfonso Rios</name>\n </author>\n <author>\n <name>Luis Marti Bonmati</name>\n </author>\n <author>\n <name>Irene Torres-Espallardo</name>\n </author>\n </entry>"
}